SOP for Calibration of Nebulizer Dosage Measurement Devices

SOP for Calibration of Nebulizer Dosage Measurement Devices

Standard Operating Procedure for Calibrating Nebulizer Dosage Measurement Devices

1) Purpose

The purpose of this SOP is to establish a standard procedure for the calibration of nebulizer dosage measurement devices to ensure accurate and consistent dosage delivery during production and testing.

2) Scope

This SOP applies to all dosage measurement devices used in the nebulizer manufacturing process, including manual and automated systems.

3) Responsibilities

Operators: Conduct preliminary checks and basic calibration procedures.
Maintenance Team: Perform advanced calibration and troubleshoot device errors.
Quality Assurance (QA): Validate the calibration results and maintain records.

4) Procedure

4.1 Pre-Calibration Preparation

  • Ensure the device is powered off before calibration.
  • Clean the device using a lint-free cloth and approved cleaning solution to remove any residue.
  • Check the device’s calibration status from the previous record and verify the next due date.
  • Gather all required tools, such as calibration weights, reference standards, and necessary documentation.
See also  SOP for Calibration of Nebulizer Pressure Relief Valves

4.2 Calibration Process

  • Turn on the device and allow it to warm up as per the manufacturer’s guidelines.
  • Connect the device to a reference standard using approved adapters and connectors.
  • Follow the steps in the device manual to access the calibration mode.
  • Input the reference standard’s values into the device system.
  • Run the calibration process and record the
output values for each test.
  • Adjust settings if the output values deviate from the reference standards.
  • Repeat the process three times to ensure consistency and accuracy.
  • 4.3 Post-Calibration Activities

    • Document the calibration results in the calibration log.
    • Label the device with the calibration date, next due date, and the initials of the person performing the calibration.
    • Perform a sample test with the calibrated device to validate its functionality.

    4.4 Frequency of Calibration

    • Monthly: Routine calibration to ensure consistent performance.
    • Quarterly: Comprehensive calibration using advanced tools and methods.
    • Annually: Third-party calibration verification (if applicable).

    5) Abbreviations

    • PPE: Personal Protective Equipment
    • QA: Quality Assurance

    6) Documents

    The following documents should be maintained:

    • Calibration Log
    • Device Validation Records
    • Error Adjustment Report

    7) References

    Relevant regulatory guidelines and references include:

    • Device Manufacturer’s Manual
    • ISO 13485: Medical Devices Quality Management Systems

    8) SOP Version

    Version: 1.0

    Annexure

    Annexure Title: Calibration Log for Nebulizer Dosage Measurement Devices

     
    Date Device Name Reference Standard Calibration Status Operator Initials Remarks
    DD/MM/YYYY Dosage Measurement Device Standard Value Pass/Fail Operator Name Details of calibration output
               

    Annexure Title: Device Validation Records

     
    Date Device Name Validation Type Operator Initials Supervisor Signature
    DD/MM/YYYY Dosage Measurement Device Routine/Comprehensive Operator Name Supervisor Name
             
    See also  SOP for Maintenance of Temperature and Humidity Control Systems in Nebulizer Production Areas

    Related Posts