Standard Operating Procedure for Calibrating Pressure Sensors in Nebulizer Systems
1) Purpose
The purpose of this SOP is to define a standardized procedure for calibrating pressure sensors used in nebulizer systems to ensure accurate and consistent pressure measurement during operation.
2) Scope
This SOP applies to all pressure sensors installed in nebulizer production equipment, including solution preparation vessels, filling machines, and assembly lines.
3) Responsibilities
Operators: Assist with sensor calibration activities and ensure equipment is prepared for calibration.
Maintenance Team: Perform the calibration of pressure sensors as per this SOP.
Quality Assurance (QA): Verify calibration results and approve calibration records.
4) Procedure
4.1 Pre-Calibration Checks
- Verify that the calibration equipment, such as pressure calibrators or manometers, is functional and within its calibration validity.
- Inspect the pressure sensor for physical damage or contamination that may affect calibration.
- Ensure the system or equipment to which the sensor is attached is clean and prepared for calibration.
- Document the pressure sensor’s details, including model number, serial number, and location, in the Calibration Log.
4.2 Calibration Procedure
- Connect the pressure calibrator to the pressure sensor using appropriate fittings and ensure a leak-free connection.
- Set the calibrator to zero and confirm the sensor reads zero pressure.
- Apply known pressure values
4.3 Post-Calibration Verification
- After calibration, repeat the pressure application and recording process to verify that the sensor is now within acceptable tolerance limits.
- Document the pre- and post-calibration readings in the Pressure Sensor Calibration Report.
- Label the sensor with the calibration date, next due date, and the initials of the person who performed the calibration.
4.4 Frequency of Calibration
- Routine Calibration: Quarterly or as per the manufacturer’s recommendation.
- Post-Maintenance Calibration: After repairs or replacement of sensor components.
- Annual Calibration: Comprehensive calibration performed by a certified external agency.
4.5 Handling Calibration Failures
- In the event of calibration failure, isolate the affected equipment and inform the QA team.
- Document the issue in the Calibration Failure Report and implement corrective actions.
- Do not use the sensor until it has been recalibrated successfully.
5) Abbreviations
- QA: Quality Assurance
- MSDS: Material Safety Data Sheet
6) Documents
The following documents should be maintained:
- Pressure Sensor Calibration Log
- Pressure Sensor Calibration Report
- Calibration Failure Report
7) References
Relevant regulatory guidelines and references include:
- ISO 13485: Medical Devices Quality Management Systems
- Good Manufacturing Practices (GMP) Guidelines
- Equipment Manufacturer’s Manual
8) SOP Version
Version: 1.0
Annexure
Annexure Title: Pressure Sensor Calibration Log
Date | Sensor ID | Equipment ID | Calibration Range | Performed By | Remarks |
---|---|---|---|---|---|
DD/MM/YYYY | Sensor Identifier | Equipment Identifier | Pressure Range | Technician Name | Details of calibration |
Annexure Title: Pressure Sensor Calibration Report
Date | Sensor ID | Applied Pressure | Sensor Reading | Deviation | Remarks |
---|---|---|---|---|---|
DD/MM/YYYY | Sensor Identifier | Pressure Applied (kPa) | Reading (kPa) | Deviation (%) | Details |
Annexure Title: Calibration Failure Report
Date | Sensor ID | Issue Identified | Corrective Action | Performed By | QA Approval |
---|---|---|---|---|---|
DD/MM/YYYY | Sensor Identifier | Description of Issue | Corrective Steps | Technician Name | QA Name |