SOP for Validation of Dissolution Testing for Tablets and Capsules

SOP for Validation of Dissolution Testing for Tablets and Capsules

Standard Operating Procedure for Validating Dissolution Testing for Tablets and Capsules

1) Purpose

The purpose of this Standard Operating Procedure (SOP) is to provide a standardized approach for validating dissolution testing methods for tablets and capsules. This SOP ensures that the dissolution test provides consistent and reliable data to evaluate the release of active pharmaceutical ingredients (APIs) from solid dosage forms, ensuring that the product meets the required specifications for bioavailability and therapeutic efficacy. The validation of key parameters such as dissolution apparatus, medium, temperature, and sampling ensures the robustness and reproducibility of the testing method, and compliance with regulatory requirements.

2) Scope

This SOP applies to the dissolution testing of tablets and capsules at the facility. It includes the validation of various dissolution apparatus (e.g., USP Apparatus 1 and Apparatus 2) and the conditions used for testing, such as dissolution medium, temperature, sampling time points, and analytical methods. The SOP is applicable to both new tablet and capsule formulations and those undergoing significant changes. It ensures that critical parameters involved in dissolution testing are validated to guarantee the accuracy, precision, and reliability of the dissolution results. This SOP is intended for use

by quality assurance (QA), analytical teams, and technical personnel involved in dissolution testing and method validation.

3) Responsibilities

Operators: Responsible for carrying out dissolution testing according to the approved protocol and ensuring that all critical process parameters are met and documented.
Quality Assurance (QA): Ensures that the dissolution testing process is validated in compliance with this SOP and meets all regulatory requirements. QA is also responsible for reviewing and approving all validation reports and documentation.
Analytical Supervisors: Oversee the dissolution testing process, ensuring that the correct methodology and parameters are followed and that the test is conducted according to the approved protocol.
Validation Team: Responsible for developing the dissolution testing validation protocol, executing the validation runs, and analyzing the results.
Maintenance Personnel: Ensures that all dissolution testing equipment, including dissolution apparatus and sampling systems, is properly calibrated and maintained during the validation process.

See also  SOP for Process Validation of Sustained-Release Capsules

4) Procedure

The following steps should be followed for the validation of dissolution testing for tablets and capsules:

1. Preparation for Validation:
1.1 Review batch records and product specifications to identify critical process parameters (CPPs) for dissolution testing, such as dissolution medium, temperature, time, and sampling intervals.
1.2 Develop a process validation protocol that includes the objectives, scope, equipment, testing method, and personnel involved in the validation process.
1.3 Ensure that all dissolution testing equipment, including dissolution vessels, paddles, baskets, and sampling systems, are calibrated and operational before initiating the validation process.

2. Validation Protocol:
2.1 Protocol Design: The validation protocol should be developed based on the formulation of the tablet or capsule and the specific dissolution apparatus used. The protocol should also outline the validation procedure for each dissolution condition.
2.2 Defining Acceptance Criteria: Establish acceptance criteria for critical parameters such as dissolution rate, release profile, and API concentration. The criteria should also include parameters for precision, accuracy, and reproducibility of the dissolution method.
2.3 Risk Assessment: Perform a risk assessment to identify critical quality attributes (CQAs) and corresponding critical process parameters (CPPs) for the dissolution testing process. This will help ensure that all risks to product quality are identified and controlled.

3. Execution of Validation:
3.1 Dissolution Apparatus Setup: Set up the dissolution apparatus according to the specified parameters (e.g., apparatus type, rotation speed, temperature) for the dissolution test.
3.2 Preparation of Dissolution Medium: Prepare the dissolution medium as per the specifications, ensuring that it is prepared in the correct volume, composition, and pH for the dissolution process.
3.3 Dissolution Testing: Conduct the dissolution test by placing the tablets or capsules in the dissolution vessel, following the predetermined time intervals for sampling. Monitor the dissolution process and ensure that it runs according to the specified conditions.
3.4 Sampling: Collect samples at the defined time points and analyze them for the amount of drug released. Ensure that the sampling process is consistent and accurate.
3.5 Data Analysis: Analyze the dissolution profiles by comparing the amount of drug released against the acceptance criteria, and ensure that the dissolution data is consistent with the predefined specifications.
3.6 Repeatability and Precision: Perform multiple runs of the dissolution test to assess the repeatability and precision of the method. Analyze the variability between different runs to confirm the robustness of the dissolution testing process.

See also  SOP for Validation of Drying Processes (Fluidized Bed and Tray Dryers)

4. Documentation and Reporting:
4.1 Record all data during the validation process, including batch records, equipment logs, process parameters, and sampling results.
4.2 Ensure that all forms, reports, and certificates are completed and signed by the responsible personnel.
4.3 Perform statistical analysis of the collected data to assess the consistency and capability of the dissolution testing process. This analysis should confirm that the dissolution process consistently produces results that meet the acceptance criteria.
4.4 Prepare a final validation report summarizing the results of the validation, including any deviations, corrective actions, and conclusions regarding the validation of the dissolution testing process.

5. Revalidation:
5.1 Revalidate the dissolution testing process if there are significant changes to the formulation, equipment, or critical process parameters.
5.2 Periodically conduct revalidation to ensure continued compliance with regulatory requirements and confirm that the dissolution testing process consistently produces accurate and reliable results.

See also  SOP for Process Validation of Injectable Dosage Forms

5) Abbreviations

  • QA: Quality Assurance
  • CPP: Critical Process Parameter
  • CQA: Critical Quality Attribute
  • SOP: Standard Operating Procedure
  • USP: United States Pharmacopeia
  • HPLC: High Performance Liquid Chromatography

6) Documents

  • Process Validation Protocol
  • Batch Production Records
  • Equipment Calibration Logs
  • Process Validation Reports
  • Sampling and Analysis Data
  • Statistical Analysis Data

7) Reference

  • United States Pharmacopeia (USP) Chapter 711: Dissolution
  • International Council for Harmonisation (ICH) Q7: Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients
  • FDA Guidance for Industry: Dissolution Testing of Immediate Release Solid Oral Dosage Forms
  • ISO 9001: Quality Management Systems – Requirements

8) SOP Version

Version 1.0 – Effective Date: DD/MM/YYYY

Annexure

Template 1: Dissolution Testing Record

Date Time Operator Initials Dissolution Medium Sample Volume (mL) Time Point (minutes)
DD/MM/YYYY HH:MM Operator Name Medium used Volume in mL Time in minutes
           

Template 2: Dissolution Profile Analysis Record

Batch No. Test Date Cumulative Release (%) Pass/Fail Operator Initials
Batch Number DD/MM/YYYY Cumulative Release in % Pass/Fail Operator Name
         

Template 3: Sampling and Analysis Record

Sample ID Test Type Test Method Result Operator Initials
Sample Number Test Name Method used Result Operator Name
         

Template 4: Calibration Log

Equipment ID Calibration Date Calibration Method Calibration Results Operator Initials
Equipment Number DD/MM/YYYY Method used Pass/Fail Operator Name
         

Template 5: Statistical Analysis Data

Sample ID Test Type Mean Standard Deviation Result
Sample Number Test Name Mean Value Standard Deviation Pass/Fail
         