Standard Operating Procedure for Calibrating Nebulizer Dosage Measurement Devices
1) Purpose
The purpose of this SOP is to establish a standard procedure for the calibration of nebulizer dosage measurement devices to ensure accurate and consistent dosage delivery during production and testing.
2) Scope
This SOP applies to all dosage measurement devices used in the nebulizer manufacturing process, including manual and automated systems.
3) Responsibilities
Operators: Conduct preliminary checks and basic calibration procedures.
Maintenance Team: Perform advanced calibration and troubleshoot device errors.
Quality Assurance (QA): Validate the calibration results and maintain records.
4) Procedure
4.1 Pre-Calibration Preparation
- Ensure the device is powered off before calibration.
- Clean the device using a lint-free cloth and approved cleaning solution to remove any residue.
- Check the device’s calibration status from the previous record and verify the next due date.
- Gather all required tools, such as calibration weights, reference standards, and necessary documentation.
4.2 Calibration Process
- Turn on the device and allow it to warm up as per the manufacturer’s guidelines.
- Connect the device to a reference standard using approved adapters and connectors.
- Follow the steps in the device manual to access the calibration mode.
- Input the reference standard’s values into the device system.
- Run the calibration process and record the
4.3 Post-Calibration Activities
- Document the calibration results in the calibration log.
- Label the device with the calibration date, next due date, and the initials of the person performing the calibration.
- Perform a sample test with the calibrated device to validate its functionality.
4.4 Frequency of Calibration
- Monthly: Routine calibration to ensure consistent performance.
- Quarterly: Comprehensive calibration using advanced tools and methods.
- Annually: Third-party calibration verification (if applicable).
5) Abbreviations
- PPE: Personal Protective Equipment
- QA: Quality Assurance
6) Documents
The following documents should be maintained:
- Calibration Log
- Device Validation Records
- Error Adjustment Report
7) References
Relevant regulatory guidelines and references include:
- Device Manufacturer’s Manual
- ISO 13485: Medical Devices Quality Management Systems
8) SOP Version
Version: 1.0
Annexure
Annexure Title: Calibration Log for Nebulizer Dosage Measurement Devices
Date | Device Name | Reference Standard | Calibration Status | Operator Initials | Remarks |
---|---|---|---|---|---|
DD/MM/YYYY | Dosage Measurement Device | Standard Value | Pass/Fail | Operator Name | Details of calibration output |
Annexure Title: Device Validation Records
Date | Device Name | Validation Type | Operator Initials | Supervisor Signature |
---|---|---|---|---|
DD/MM/YYYY | Dosage Measurement Device | Routine/Comprehensive | Operator Name | Supervisor Name |