SOP for Calibration of Nebulizer Temperature Sensors

SOP for Calibration of Nebulizer Temperature Sensors

Standard Operating Procedure for Calibrating Nebulizer Temperature Sensors

1) Purpose

The purpose of this SOP is to provide a standardized method for calibrating temperature sensors used in nebulizer manufacturing to ensure accurate temperature readings and compliance with quality standards.

2) Scope

This SOP applies to all temperature sensors utilized in nebulizer production equipment, including mixing tanks, storage units, and sterilization chambers, within the facility.

3) Responsibilities

Operators: Assist with calibration setup and documentation.
Maintenance Team: Perform calibration and ensure proper functioning of temperature sensors.
Quality Assurance (QA): Validate calibration results and approve calibration records.
Supervisors: Ensure adherence to this SOP during calibration activities.

4) Procedure

4.1 Pre-Calibration Checks

  • Ensure the calibration equipment, such as standard thermometers or temperature calibrators, is within its calibration validity period.
  • Inspect the temperature sensor for visible damage, dirt, or corrosion.
  • Verify that the sensor is properly connected to the equipment and power supply.
  • Document the sensor details, including model, serial number, and location, in the Calibration Log.
See also  SOP for Cleaning and Maintenance of Nebulizer Dispensing Systems

4.2 Calibration Setup

  • Disconnect the temperature sensor from the production equipment if required.
  • Connect the sensor to a certified temperature calibrator or immerse it in a temperature-controlled bath.
  • Set the calibrator or bath to a known reference temperature.
  • Allow the sensor to stabilize at the reference temperature for a specified duration.

4.3 Calibration Procedure

  • Record the temperature reading displayed by the sensor and compare it with the reference temperature.
  • Repeat the measurement at multiple reference temperatures within the sensor’s operating range (e.g., 0°C, 25°C, 50°C).
  • Calculate the deviation for each measurement and ensure it is within the acceptable range specified by the manufacturer.
  • If deviations exceed the acceptable limits, adjust the sensor settings or recalibrate using the manufacturer’s instructions.

4.4 Post-Calibration Activities

  • Reinstall the temperature sensor onto the production equipment, ensuring proper connections.
  • Label the sensor with the calibration status, date, and next due date.
  • Submit the calibration results and supporting documents to QA for validation and approval.
  • Record the calibration activity in the Calibration Log and retain the Calibration Report.
See also  SOP for Operation of Nebulizer Component Assembly Machines

4.5 Frequency of Calibration

  • Routine Calibration: Semi-annually or as per the manufacturer’s recommendations.
  • Post-Maintenance Calibration: After repairs or replacement of critical components.
  • Annual Calibration: Comprehensive calibration by an external certified agency.

4.6 Handling Calibration Failures

  • Isolate the equipment if the sensor fails calibration and report the issue to the maintenance team.
  • Document the failure details in the Calibration Failure Report and notify QA.
  • Replace or repair the faulty sensor and repeat the calibration process before resuming production.

4.7 Safety Precautions

  • Wear appropriate personal protective equipment (PPE), including gloves and safety goggles, during calibration activities.
  • Handle calibration equipment and temperature sensors carefully to avoid damage.
  • Ensure all electrical connections are secure and follow lockout/tagout procedures if necessary.
See also  SOP for Maintenance of Nebulizer Conveyor Belts

5) Abbreviations

  • QA: Quality Assurance
  • PPE: Personal Protective Equipment

6) Documents

The following documents should be maintained:

  • Calibration Log
  • Calibration Report
  • Calibration Failure Report

7) References

Relevant regulatory guidelines and references include:

  • ISO 13485: Medical Devices Quality Management Systems
  • Good Manufacturing Practices (GMP) Guidelines
  • Equipment Manufacturer’s Manual

8) SOP Version

Version: 1.0

Annexure

Annexure Title: Calibration Log

 
Date Sensor ID Reference Temperature Measured Temperature Deviation (%) Performed By
DD/MM/YYYY Sensor Identifier Reference Value Measured Value Deviation Value Technician Name
           

Annexure Title: Calibration Report

 
Date Sensor ID Pre-Calibration Reading Post-Calibration Reading Performed By QA Approval
DD/MM/YYYY Sensor Identifier Details of pre-readings Details of post-readings Technician Name QA Name
           

Annexure Title: Calibration Failure Report

 
Date Sensor ID Issue Identified Corrective Action Performed By QA Approval
DD/MM/YYYY Sensor Identifier Details of issue Details of corrective action Technician Name QA Name
           

Related Posts