SOP for pH Determination in Oral Liquids

SOP for pH Determination in Oral Liquids

Procedure for pH Determination in Oral Liquids

1) Purpose

The purpose of this SOP is to outline the procedure for determining the pH of oral liquid pharmaceutical products to ensure compliance with specifications and regulatory requirements.

2) Scope

This SOP applies to all oral liquid pharmaceutical products requiring pH determination to assess acidity or alkalinity as part of quality control measures.

3) Responsibilities

Quality Control (QC) Department: Responsible for performing pH determination tests.
Quality Assurance (QA) Department: Responsible for review and approval of pH testing procedures and results.
Manufacturing Department: Responsible for providing oral liquid samples and support during testing.

4) Procedure

4.1 Equipment Preparation:
    4.1.1 Calibrate the pH meter according to manufacturer’s instructions.
    4.1.2 Ensure electrodes are clean and conditioned prior to testing.

See also  SOP for Bulk Density and Tapped Density Testing

4.2 Sample Preparation:
    4.2.1 Mix the oral liquid sample thoroughly to ensure homogeneity.
    4.2.2 Transfer a representative sample into a clean, dry beaker.

4.3 pH Measurement:
    4.3.1 Immerse the pH electrode into the oral liquid sample without touching the bottom or sides of the beaker.
    4.3.2 Allow the pH reading to stabilize as per meter instructions, then record the pH value.

4.4 Calibration Verification:
    4.4.1 Verify pH meter calibration periodically during testing with standard buffer solutions.
    4.4.2 Adjust pH meter if readings deviate from expected values.

4.5 Data Recording:
    4.5.1 Record the pH value obtained from each sample tested.
    4.5.2 Document any observations or deviations from expected results.

See also  SOP for Spreadability Testing in Semisolids

4.6 Reproducibility Testing:
    4.6.1 Repeat pH measurements on the same sample to confirm reproducibility.
    4.6.2 Compare results and ensure consistency within acceptable limits.

4.7 Data Analysis:
    4.7.1 Compile and review pH measurement data, including initial readings, calibration checks, and reproducibility testing results.
    4.7.2 Analyze data against acceptance criteria and specifications.

4.8 Reporting:
    4.8.1 Prepare a pH Determination Report summarizing test procedures, results, and conclusions.
    4.8.2 Include any corrective actions or recommendations based on pH measurement findings.
    4.8.3 Submit the report to QA for review and approval and maintain records as per facility procedures.

5) Abbreviations, if any

QC: Quality Control
QA: Quality Assurance

See also  SOP for Optical Rotation Testing

6) Documents, if any

pH Determination Report
Calibration Records
Sample Testing Data

7) Reference, if any

USP (insert appropriate reference for pH testing)
FDA Guidance for Industry – Container Closure Systems for Packaging Human Drugs and Biologics

8) SOP Version

Version 1.0