Standard Operating Procedure for Preparing Nasal Sprays for Systemic Drug Delivery
1) Purpose
This SOP describes the procedure for formulating and preparing nasal sprays intended for systemic drug delivery. These formulations are designed to ensure the rapid absorption of drugs through the nasal mucosa and into the bloodstream.
2) Scope
This SOP applies to personnel involved in the preparation, testing, and quality control of nasal sprays designed for systemic absorption of drugs at [Company Name]. It includes procedures for the preparation, homogeneity testing, and packaging of systemic nasal sprays.
3) Responsibilities
- Operators: Responsible for accurately weighing, blending, and filling the nasal spray formulations for systemic drug delivery.
- Quality Assurance (QA): Ensures that the formulation meets required systemic absorption profiles and regulatory standards.
- Maintenance Team: Ensures the calibration and cleaning of all equipment involved in the preparation and filling of systemic nasal sprays.
4) Procedure
4.1 Preparation of Formulation
4.1.1 Selection of Drug and Excipients
- Select drugs that are known to have high bioavailability through the nasal route. Ensure the drug is pharmaceutically acceptable and is suitable for systemic delivery.
- Choose excipients that enhance absorption (e.g., surfactants, permeation enhancers) and maintain the stability of the drug. Record all components in the batch manufacturing record (BMR).
4.1.2 Weighing and Mixing
- Weigh the required amounts of drug and excipients using a calibrated balance. Record the weights in the BMR.
- Add the drug and excipients to the mixing vessel. Use a magnetic stirrer or mechanical mixer to ensure uniform distribution of the drug within the excipients.
4.1.3 Homogenization
- If necessary, use a homogenizer to achieve a consistent particle size and uniform drug distribution. Homogenize the mixture at the recommended speed and duration as specified in the BMR.
- Document the homogenization parameters, including speed and time, in the BMR.
4.2 Filling and Packaging
4.2.1 Filling Process
- Transfer the homogenized formulation into the nasal spray filling machine. Set the machine to fill the appropriate volume for each spray dose (e.g., 50 µL per spray).
- Test the fill accuracy by checking 10 randomly filled units. The variation should not exceed ±5% of the target dose. Record the results in the BMR.
4.2.2 Sealing and Packaging
- Seal the filled nasal spray containers with sterile caps or nozzles. Ensure that the spray mechanism is functioning correctly by testing one or two units.
- Package the sealed units in protective containers and label each unit with the batch number, expiration date, and storage instructions.
4.3 Quality Control
4.3.1 Homogeneity Testing
- Test the homogeneity of the nasal spray formulation by sampling from different points in the mixing vessel. Use HPLC or another validated analytical technique to measure drug concentration in each sample.
- Ensure the drug concentration across samples does not vary by more than 2%. Document the results in the homogeneity testing log.
4.3.2 pH and Osmolality Testing
- Test the p
You said:
Continue
ChatGPT said:
ChatGPT
html
Copy code
h of the nasal spray to ensure it falls within the acceptable range for nasal administration (typically 4.5-7.0). - Test the osmolality to ensure the formulation is isotonic or near-isotonic with nasal fluids (approximately 270-330 mOsm/kg). Document all results in the pH and osmolality testing log.
4.3.3 Absorption Testing
- Perform in vitro or in vivo absorption testing to ensure the drug is rapidly absorbed through the nasal mucosa. Use validated analytical methods to measure the rate and extent of absorption.
- Record the results of the absorption testing in the batch manufacturing record (BMR) and compare them to target specifications for systemic delivery.
4.4 Stability Testing
4.4.1 Long-Term Stability Testing
- Conduct long-term stability testing on the formulation, storing samples at controlled temperatures (e.g., 25°C, 60% RH) and testing them at 1-month, 3-month, and 6-month intervals.
- Measure the physical and chemical properties, including drug concentration, pH, and osmolality. Document all results in the stability testing log.
4.4.2 Accelerated Stability Testing
- Conduct accelerated stability testing by storing samples at elevated temperatures (e.g., 40°C, 75% RH) to predict the shelf life of the formulation. Test samples at specified intervals.
- Document all results in the stability testing log and adjust the product expiration date accordingly.
4.5 Documentation
- Document all steps of the formulation and preparation process in the batch manufacturing record (BMR), including ingredient weights, mixing times, test results, and filling parameters.
- Ensure QA personnel review and sign off on all documents before the product is released for distribution.
4.6 Equipment Cleaning and Calibration
- After completing the batch, clean and sterilize all equipment, including mixers and filling machines, following the cleaning validation procedure. Document all cleaning activities in the equipment cleaning log.
- Calibrate the equipment as per the calibration schedule, and record the results in the calibration log.
5) Abbreviations, if any
- QA: Quality Assurance
- HPLC: High-Performance Liquid Chromatography
- BMR: Batch Manufacturing Record
6) Documents, if any
- Batch Manufacturing Record (BMR)
- Homogeneity Testing Log
- pH and Osmolality Testing Log
- Stability Testing Log
- Absorption Testing Log
- Cleaning Log
- Calibration Log
7) References, if any
- ICH Q1A – Stability Testing Guidelines
- FDA Guidance for Nasal Drug Products
8) SOP Version
Version 1.0
Annexure
1. Homogeneity Testing Log Template
Date | Formulation | Test Point (Top/Middle/Bottom) | Drug Concentration (%) | Operator Initials | QA Approval |
---|---|---|---|---|---|
DD/MM/YYYY | Formulation Name | Test Point | Concentration | Operator Name | QA Name |
2. pH and Osmolality Testing Log Template
Date | Formulation | pH | Osmolality (mOsm/kg) | Operator Initials | QA Approval |
---|---|---|---|---|---|
DD/MM/YYYY | Formulation Name | pH Value | Osmolality | Operator Name | QA Name |
3. Stability Testing Log Template
Date | Formulation | Storage Conditions | Time Interval | Stability Results | Operator Initials | QA Approval |
---|---|---|---|---|---|---|
DD/MM/YYYY | Formulation Name | Temperature/Humidity | 1 month, 3 months, etc. | Pass/Fail | Operator Name | QA Name |
4. Absorption Testing Log Template
Date | Formulation | Absorption Test Method | Drug Absorption Rate | Operator Initials | QA Approval |
---|---|---|---|---|---|
DD/MM/YYYY | Formulation Name | Test Method (e.g., In Vitro, In Vivo) | Absorption Rate | Operator Name | QA Name |
5. Cleaning and Calibration Log Template
Date | Equipment ID | Cleaning/Calibration Procedure | Operator Initials | QA Approval | |
---|---|---|---|---|---|
DD/MM/YYYY | Equipment Name/ID | Procedure | Operator Name | QA Name | |