SOP for Sterilization Process Validation

Posted on By

SOP for Sterilization Process Validation

Comprehensive Guide to Sterilization Process Validation in Medical Device Manufacturing

1) Purpose

The purpose of this SOP is to establish a standardized process for validating sterilization procedures used in medical device manufacturing. Validation ensures that sterilization processes consistently achieve the required sterility assurance level (SAL) and comply with regulatory and quality standards.

2) Scope

This SOP applies to all sterilization processes, including steam, ethylene oxide (EO), dry heat, and radiation methods, used in the production of medical devices. It covers initial validation, periodic revalidation, and monitoring. The SOP is intended for use by validation teams, quality assurance personnel, and production staff.

3) Responsibilities

– Validation Team: Plans, executes, and documents sterilization validation activities.
– Quality Assurance (QA): Reviews and approves validation protocols and results.
– Production Team: Provides operational input and ensures adherence to validated procedures.
– Maintenance Team: Ensures sterilization equipment is properly maintained and calibrated.

4) Procedure

4.1 Planning and Preparation
4.1.1 Develop Validation Plan
– Create a Sterilization Validation Master Plan (SVMP) outlining the scope, objectives, and responsibilities.
– Define acceptance criteria for sterilization process performance, including SAL, temperature, humidity, and exposure time.
4.1.2 Protocol Preparation
– Develop specific validation protocols for installation qualification (IQ), operational qualification (OQ), and performance qualification

(PQ).
– Include biological and chemical indicators, load configurations, and sampling plans.

See also  SOP for Medical Device Assembly Procedures

4.2 Installation Qualification (IQ)
4.2.1 Objective
– Ensure sterilization equipment is installed as per manufacturer specifications and operates correctly.
4.2.2 Activities
– Verify utility connections (e.g., steam, electricity) and control systems.
– Confirm that monitoring instruments (e.g., thermocouples, pressure gauges) are calibrated.
– Document installation details, including equipment model and serial numbers.

4.3 Operational Qualification (OQ)
4.3.1 Objective
– Validate that the sterilization equipment operates within specified parameters.
4.3.2 Activities
– Conduct empty chamber heat distribution studies to confirm uniform temperature and pressure.
– Test the equipment under varying load conditions to identify potential inconsistencies.
– Document results and address deviations.

4.4 Performance Qualification (PQ)
4.4.1 Objective
– Verify that the sterilization process consistently achieves the desired SAL for actual production loads.
4.4.2 Activities
– Run full production loads with biological and chemical indicators placed in the most challenging locations.
– Measure and record key parameters, such as temperature, pressure, and exposure time.
– Analyze sterility test results to confirm effectiveness.
4.4.3 Documentation
– Record load configurations, test results, and corrective actions (if any) in the PQ report.

See also  SOP for Supplier Qualification and Audits

4.5 Revalidation
4.5.1 Schedule
– Perform periodic revalidation (e.g., annually) or after significant equipment/process changes.
4.5.2 Activities
– Repeat OQ and PQ steps to confirm ongoing compliance.
– Update validation documentation as necessary.

4.6 Monitoring and Controls
4.6.1 Routine Monitoring
– Monitor sterilization cycles in real-time using automated systems.
– Use biological and chemical indicators for batch-level verification.
4.6.2 Process Deviations
– Investigate any deviations from validated parameters.
– Halt production if sterility cannot be guaranteed and implement corrective actions.

4.7 Documentation and Record Keeping
4.7.1 Validation Reports
– Compile detailed reports summarizing IQ, OQ, and PQ activities, results, and conclusions.
4.7.2 Calibration Records
– Maintain calibration certificates for monitoring instruments.
4.7.3 Change Control
– Document all changes to the sterilization process, including updates to equipment or protocols.

See also  SOP for Production Planning and Scheduling

5) Abbreviations

– SAL: Sterility Assurance Level
– IQ: Installation Qualification
– OQ: Operational Qualification
– PQ: Performance Qualification
– SVMP: Sterilization Validation Master Plan

6) Documents

– Sterilization Validation Master Plan
– IQ, OQ, and PQ Protocols
– Calibration Certificates
– Sterilization Cycle Records
– Validation Reports

7) Reference

– ISO 13485: Medical devices – Quality management systems
– ISO 11135: Sterilization of health care products – Ethylene oxide
– ISO 11137: Sterilization of health care products – Radiation
– FDA CFR Title 21, Part 820: Quality System Regulation

8) SOP Version

– Version: 1.0
– Effective Date: DD/MM/YYYY
– Approved by: [Name/Title]

Annexure

Annexure 1: Sterilization Cycle Log Template

Cycle Date Equipment ID Load Configuration SAL Achieved Operator Initials Remarks
DD/MM/YYYY Equipment ID Configuration Details 10^-6 Operator Name Comments

Annexure 2: Validation Test Summary Template

Test Parameter Setpoint Actual Value Pass/Fail Remarks
Temperature 121°C 121.2°C Pass Within Tolerance
Medical Devices Tags:FDA Medical device SOP requirements, GMP SOP for medical devices, ISO 13485 SOP for Medical device, Medical device compliance SOP, Medical device packaging SOP template, Medical device post-market surveillance SOP, Medical device quality control SOP, Medical device quality SOP, Medical device regulatory SOP, Medical device SOP, SOP for change control in medical device, SOP for device design control, SOP for equipment calibration in medical devices, SOP for handling complaints in medical devices, SOP for manufacturing medical devices, SOP for medical device audits, SOP for medical device inspection, SOP for medical device labeling, SOP for medical device manufacturing, SOP for medical device packaging inspection, SOP for Medical device raw material, SOP for medical device testing, SOP for non-conforming products in medical devices, SOP for packaging medical devices, SOP for product validation, SOP for production planning in medical devices, SOP for risk management in medical devices, SOP for sterilization packaging validation, SOP for sterilization process of Medical device, SOP for traceability in medical device manufacturing

Post navigation

Previous Post: Ointments: SOP for Cleaning Homogenizers – V 2.0
Next Post: SOP for Screening Drug Nanoparticles for Solubility Enhancement

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

English
 - 
en
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2025 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version